background: Antibodies against factor (F)VIII are a major complication in the treatment of patients with severe hemophilia A. The Nijmegen-Bethesda assay (NBA) is the gold standard for detection of neutralizing antibodies (inhibitors), whereas both inhibitors and nonneutralizing antibodies can be detected by immunoassays such as enzyme-linked immunosorbent assay (ELISA) and multiplex bead-based assays.
objective: Evaluation of an in-house Luminex bead-based assay (LumiTope) compared with a commercially available ELISA and NBA.
methods: The LumiTope method comprised full-length and B-domain-deleted FVIII as well as 9 purified FVIII single or multidomains. The respective proteins were coupled to magnetic beads to... More
background: Antibodies against factor (F)VIII are a major complication in the treatment of patients with severe hemophilia A. The Nijmegen-Bethesda assay (NBA) is the gold standard for detection of neutralizing antibodies (inhibitors), whereas both inhibitors and nonneutralizing antibodies can be detected by immunoassays such as enzyme-linked immunosorbent assay (ELISA) and multiplex bead-based assays.
objective: Evaluation of an in-house Luminex bead-based assay (LumiTope) compared with a commercially available ELISA and NBA.
methods: The LumiTope method comprised full-length and B-domain-deleted FVIII as well as 9 purified FVIII single or multidomains. The respective proteins were coupled to magnetic beads to detect domain-specific immunoglobulin (IgG; IgG) anti-FVIII antibodies in a large cohort of patients with hemophilia A with and without inhibitors.
results: Overall, LumiTope assay had a high sensitivity (94.9%) and specificity (91.2%), particularly in patients with low-titer inhibitors compared with ELISA (sensitivity of 72.2% vs 27.7%). IgG was the most abundant IgG subclass in NBA-positive patients. NBA-positive and -negative patients showed different domain profiles. Patients with genetic variants in the heavy chain predominantly exhibited antibodies specific to this chain, while those with a light-chain variant showed a more diverse distribution of antibody specificities. Patients with an intron 22 inversion resembled those with a light-chain defect, with a majority of antibodies targeting the light chain.
conclusions: LumiTope assay provides a sensitive and specific method for not only detection but also domain specification of anti-FVIII-antibodies. Implementation of bead-based assays could improve antibody detection, profiling, and comparability of results and complement NBA.